A Guard Band Changes the Acceptance Boundary Used for a Decision

The specified product limit and the acceptance boundary used by a decision rule need not be identical. If a guard band is used, its purpose and basis should be explicit.

ILAC's 2019 decision-rule guidance discusses guard bands and acceptance limits. ILAC — Decision Rules and Statements of Conformity (2019).

Ask which boundary applies

For a laboratory statement, distinguish the actual requirement from the rule used to evaluate the measured result. Do not silently rewrite the product specification when explaining the decision.

Topic map for A Guard Band Changes the Acceptance Boundary Used for a Decision: Specification boundary; Acceptance boundary; Decision risk.
Topic map: Specification boundary → Acceptance boundary → Decision risk.

Keep the basis documented

Use the applicable agreed method and competent analysis. No universal guard-band size is proposed here. The relevant question is whether the rule and its risk treatment fit the intended conformity statement and are understandable to the recipient of the report.

Review distinctions for A Guard Band Changes the Acceptance Boundary Used for a Decision: Uncertainty; Guard band; Rule.
Review distinctions: Uncertainty; Guard band; Rule.